Company Products Dealers Contact

<< back to product list

 3/30



 

Main features:
VC-CMV, PC-CMV
VC-IMV, PC-IMV
oxygen therapy
- durable case
- long lasting battery

3/30, a portable ventilator designed for ambulance and rescue services, pre-hospital and hospital care. It has proved itself as a straightforward and reliable tool for conduction of volume or pressure controlled mandatory and intermittent ventilation and oxygen therapy (inhalation), with possibility to adjust inspiration attempt duration.

 

Technical data
Dimensions 210 x 160 x 150 mm
Weight3 kg (w/o cylinder or reducer)
-
Features
Ventilation modesVC-CMV, PC-CMV, VC-IMV, PC-IMV, oxygen therapy
ControlsVentilation mode, tidal volume, respiratory rate, minute volume, O2 concentration, inspiration pause duration
OptionsInvasive/ Non-Invasive (mask) ventilation, modes for Adults / Children , Wall-mountable (e.g. in ambulance)
-
Operating data
Supply gasMedical Oxygen
Supply pressure0.2 - 0.6 MPa
O2 concentration60% mix, 100%
Ventilation respiratory rate10 to 40 1/min ± 5% (ventilation modes)
Tidal volume Vt100 to 2000 mL. BTPS
Minute ventilation1 to 15 L/min (ventilation modes), 1 to 40 L/min (oxygen therapy)
I:E ratio1:2
Inspiraton time/Pause0.25 to 6 sec
PEEP0 to 20 mbar/ cmH2O
Pressure limit 15-50 cmH2O
-
Acoustic / visual
warning signals
Supply pressure low When supply pressure < 5 mbar / cmH2O
Airway pressure highWhen set pressure is exceeded
Airway pressure low / leakageWhen set pressure is low / not recovered after 1.5 minutes
High respiratory ratePatient breathes at high spotaneous rate
No inhalation warningLack of inhalation attempt from patient
Inhalation attemptInhalation attempt counter (1-5)
Low battery warning10% and less
-
Power supply
Input voltage15 V ± 0.5 V DC
AC/DC power packInput: 220 ± 22 V AC, Output: 19 V DC
Automobile DC connector9.5 - 30 V DC, Output: 19 V DC
Battery typeReplaceable Lithium-ion battery 2200mAh with automatic charger
Operating timeApproximately 6 hours (fully charged, typical ventilation)
Battery charging timeMax. 5 hours
-
Safe operating conditions
TemperatureTemperature -15 to 40 °C / 0 to 105 °F
Relative humidity5 to 95%
-
Warranty1 year (up to 36 months)
Certificates

 Certificate of Conformity No. FSR 2009/06038

 Registration Certificate No. ROSS RU.ИМ25.ВО2955

 CE certificate coming soon


Video

Video demonstration not yet implemented in English.


   Download (PDF)